Skip to content

Assay Performance Data

1. Consolidated Performance Table

Assay LOD LOQ Linear Range Intra-assay CV% Inter-assay CV%
CRP (canine) 0.5 mg/L 1.0 mg/L 1–300 mg/L < 8% < 10%
SAA (feline) 0.5 µg/mL 1.0 µg/mL 1–250 µg/mL < 8% < 10%
SDMA 0.5 µg/dL 1.0 µg/dL 1–100 µg/dL < 8% < 10%
NT-proBNP 15 pmol/L 30 pmol/L 30–10,000 pmol/L < 8% < 10%
T4 0.1 µg/dL 0.2 µg/dL 0.2–10 µg/dL < 8% < 10%
TSH 0.03 ng/mL 0.05 ng/mL 0.05–5 ng/mL < 8% < 10%
Cortisol 0.2 µg/dL 0.5 µg/dL 0.5–30 µg/dL < 8% < 10%
Progesterone 0.2 ng/mL 0.5 ng/mL 0.5–40 ng/mL < 8% < 10%
Relaxin 0.1 ng/mL 0.3 ng/mL 0.3–20 ng/mL < 8% < 10%

2. Definitions

Metric Definition
LOD Limit of Detection — lowest concentration distinguishable from blank (mean blank + 3 SD)
LOQ Limit of Quantitation — lowest concentration measurable with CV% ≤ 20%
Linear range Concentration range where signal is proportional (r² ≥ 0.99)
CV% Coefficient of variation — (SD / mean) × 100

3. Method

All assays use fluorescence immunochromatography (FICT) on the FIA680/FIA880 platform. Validation methodology follows the standard panel documented in the research center.

For per-biomarker data cards, see Biomarker Data.

Authored by: Migibio Clinical & Scientific Affairs, Guangzhou Magic Biotech Co., Ltd.

Reviewed by: Migibio R&D Quality Committee

Last updated: 2026-08-13

Disclosure: Migibio (Guangzhou Magic Biotech Co., Ltd.) is the manufacturer of the FIA680/FIA880 analyzers and FICT reagents referenced in this content. See our Editorial & Review Policy.

Contact: Martin.Wong  ·  Phone (WhatsApp): +86 13323237275  ·  Email: [email protected]